Experiment 3 Fatty Acid Determination Using Gas Chromatography



CHM510
ANALYTICAL SEPARATION METHOD

EXPERIMENT 3:
FATTY ACID DETERMINATION USING GAS CHROMATOGRAPHY (GC)


INTRODUCTION

Gas chromatography separates the analytes that is volatile and chemically stable. Fatty acids are not sufficiently volatile for GC analysis, so that it needs to be modified chemically to produce a new compound which has properties that are suitable for analysis. If the unsuitable sample is introduced into GC analysis, it tends to cause peak tailing due to the adsorption and non-specific interaction with the column. In this experiment, the fatty acid is changed to fatty acid methyl ester (FAME) that is more volatile, suitable for GC analysis by using esterification reaction that used methanolic solution with catalyst of esterification reagent. The objective for this experiment is to introduce a derivatization procedure routinely used for fat analysis in which non-volatile fatty acids are chemically converted to the corresponding volatile methyl ester (FAME) and to determine the amount of FAME in the derivatives samples. The amount of FAME is determined by using the response factor calculation:




PROCEDURE


a)      Preparation of fatty acid methyl ester samples from fat samples.

2 g of oil or fat was weighed out and the exact weight was recorded. The sample was transferred into a 50 mL flask equipped with air condenser. Then, 5 mL of 0.5 M methanolic solution was added and refluxed for 3-4 minutes. 15 mL of esterification reagent was added and refluxed about 3 minutes. After that, the mixture was transferred into a separatory flask. 50 mL of saturated sodium chloride and 25 mL of diethyl ether was added. The mixture was shaking vigorously for 2 minutes and the aqueous layer was discarded. In addition, 25 mL of saturated sodium chloride was added and the aqueous layer was discarded. The organic layer was transferred into a screw cap vial. The samples were sealed by parafilm for the quantitative analysis.



b)     Instrument set-up
Injection Port: Split (40:1)
Injection Port Temperature: 250oC
Column Temperature: 100oC to 290oC at 40oC min-1
Carrier Gas Flow Rate: 30 mL s-1
Detector Temperature: 250oC



c)      Quantitative analysis of FAME
0.4 µL of standard esters was injected onto the column. Injection was repeated to get reproducible peak areas. 0.4 µL of derivative sample was injected. Injection was repeated to get reproducible peak areas. The amount of each fatty acid in the sample was calculated using the data from the standard esters.



RESULTS


A.     Response Factor (RF) for analytes in standard FAME:

Peak
Amount of FAME in standard (ppm)
Peak Area(pA*s)
Response Factor
2
100
69.50613
1.4387
3
100
321.35806
0.31112
4
100
560.65057
0.17836
5
100
61.21711
1.6335
6
100
487.67694
0.2051






 Comparison of retention time for standard and samples:
Peak
Retention time for standard (min)
Retention time for sample 1 (min)
Retention time for sample 2 (min)
Retention time for sample 3 (min)
2
1.981
2.014
2.016
2.015
3
2.578
2.660
2.652
2.653
4
3.605
3.751
3.756
3.755
5
5.314
5.669
5.702
5.673
6
5.902
6.137
6.144
6.140




C.      Amount of FAME in samples:

Sample
Peak
Response factor of corresponding peak
Peak area (pA*s)
Amount of Fame (ppm)
1
2
1.4387
30.92022
44.48
3
0.31112
2.56153
0.80
4
0.17836
54.58322
9.74
5
1.6335
34.99018
57.16
6
0.2051
9.04965
1.86
2
2
1.4387
17.52389
25.21
3
0.31112
33.56306
10.44
4
0.17836
1261.23462
224.95
5
1.6335
1068.38159
1745.20
6
0.2051
263.29333
54.00
3
2
1.4387
5.08862
7.32
3
0.31112
21.01284
6.54
4
0.17836
848.58557
151.35
5
1.6335
1.15695
1.89
6
0.2051
180.49922
37.02






DISCUSSION

The components in the samples are compared with the standard components by the retention time. From the retention time of standard and samples, it is proven that component 5 (peak 6) is not present in all 3 samples because of the difference of the retention time between standard and samples are too big. The amount of each component is different in each samples may due to the different mass of the fat initially. Peak 5 in each sample give very large different in the amount of FAME, this may be due to the un-complete separation process during shaking process or the discarding process. It is necessary to discard little organic layer to make sure that there is no aqueous layer anymore to be injected onto GC. That condition also affected by the contaminants in the flask that is not clean before using. The other peaks that correspond to specific component show small difference that assumed to be correct. There is another way to derivatize or modified the low volatility fatty acid to more volatile compound called silylation that BSTFA to yield trimethylsilyl (TMS) ester that is more suitable to be analysed in GC.




CONCLUSION

The derivatization technique used in this experiment is esterification to convert non-volatile fatty acids to more volatile fatty acid methyl ester (FAME). There are 5 components in the standard mixture while the 3 samples only indicate 4 components as shown in the standard mixture by comparison of the retention time. The concentration of each component is calculated by using the response factor of the standard.


REFERENCE

1.      Nor’ashikin S., Ruziyati T., Mardiana S. (2012), Analytical Separation Methods Laboratory Guide (2nd edition).
2.      https://www.ncbi.nlm.nih.gov/pubmed/85206893.    http://www.doiserbia.nb.rs/img/doi/0352-5139/2015/0352-51391400073W.pdf






Artikel ini hanyalah simpanan cache dari url asal penulis yang berkebarangkalian sudah terlalu lama atau sudah dibuang :

http://fareastzareul1994.blogspot.com/2018/05/fatty-acid-determination-using-gas.html

Kempen Promosi dan Iklan
Kami memerlukan jasa baik anda untuk menyokong kempen pengiklanan dalam website kami. Serba sedikit anda telah membantu kami untuk mengekalkan servis percuma aggregating ini kepada semua.

Anda juga boleh memberikan sumbangan anda kepada kami dengan menghubungi kami di sini
Experiment 1 Gas Chromatoghraphy Optimization Of Flow Rate And Column Temperature

Experiment 1 Gas Chromatoghraphy Optimization Of Flow Rate And Column Temperature

papar berkaitan - pada 26/2/2020 - jumlah : 162 hits
CHM510ANALYTICAL SEPARATION METHODEXPERIMENT 1 GAS CHROMATOGHRAPHY OPTIMIZATION OF FLOW RATE AND COLUMN TEMPERATUREABSTRACTThe separation of the compounds will travel faster through the column when column temperature is high volatility of c...
Experiment 2 High Performance Liquid Chromatography Method Development

Experiment 2 High Performance Liquid Chromatography Method Development

papar berkaitan - pada 26/2/2020 - jumlah : 136 hits
CHM510ANALYTICAL SEPARATION METHODEXPERIMENT 2 HIGH PERFORMANCE LIQUID CHROMATOGRAPHY METHOD DEVELOPMENTOBJECTIVETo optimize a separation of a mixture of 5 compounds using HPLC by verifying the mobile phase composition INTRODUCTIONHigh Perf...
Pgn Perpanjang Pembelian Gas Dari Ophir Indonesia

Pgn Perpanjang Pembelian Gas Dari Ophir Indonesia

papar berkaitan - pada 25/2/2020 - jumlah : 129 hits
Sejak 3 tahun ini sebagai upaya untuk memenuhi kebutuhan gas yang meningkat PGN memperpanjang pasokan gas di Jawa Timur dengan PT Ophir Indonesia Pty Ltd
Medco Dan Pertamina Ep Serahkan Pipa Penyaluran Gas Di Tarakan Ke Negara

Medco Dan Pertamina Ep Serahkan Pipa Penyaluran Gas Di Tarakan Ke Negara

papar berkaitan - pada 22/2/2020 - jumlah : 183 hits
Deputi Pengendalian Pengadaan SKK Migas Tinggal mengatakan ini menunjukkan komitmennya untuk mendukung upaya optimalisasi pemanfaatan aset yang digagas oleh SKK Migas
Harga Gas Indonesia Dinilai Masih Kompetitif

Harga Gas Indonesia Dinilai Masih Kompetitif

papar berkaitan - pada 26/2/2020 - jumlah : 162 hits
Harga gas yang dipatok pemerintah USD 6 per MMBTU yang rencananya akan ditetapkan adalah belum terlalu detail kajiannya sebab unsur yang mendukung angka tersebut belum menjelaskan dampak dampak positif yang timbul
Bos Pertamina Minta Insentif Untuk Turunkan Harga Gas

Bos Pertamina Minta Insentif Untuk Turunkan Harga Gas

papar berkaitan - pada 26/2/2020 - jumlah : 168 hits
Adapun sejumlah insentif yang dimaksud antara lain tambahan bagian kontraktor produsen gas insentif pajak dan relaksasi perlakuan aset negara
Kesal Utang Ditagih Ibu Rumah Tangga Di Bandung Hantam Rentenir Pakai Tabung Gas

Kesal Utang Ditagih Ibu Rumah Tangga Di Bandung Hantam Rentenir Pakai Tabung Gas

papar berkaitan - pada 19/2/2020 - jumlah : 204 hits
Saat itu Taruli yang diduga seorang rentenir menagih hutang dengan cara berteriak teriak Tidak terima dengan perlakuan itu S langsung membawa tabung gas di dapur dan menghantamkannya sebanyak empat kali ke Taruli
Pemerintah Harus Gelontorkan Subsidi Agar Harga Gas Bisa Turun

Pemerintah Harus Gelontorkan Subsidi Agar Harga Gas Bisa Turun

papar berkaitan - pada 19/2/2020 - jumlah : 195 hits
Anggota Komisi VII DPR Ridwan Hisjam mengatakan harus ada subsidi dari pemerintah agar harga gas bisa turun menjadi USD6 per MMBTU Hingga kini harga gas bumi dari sisi hulu sekitar USD8 hingga USD9 per MMBTU
Rasa Masakan Korea Soondubu Jigae Beef Di Dubuyo

China S President To Host Official Dinner For Anwar

Drama One Cent Thief 2 Lakonan Syafiq Kyle Azira Shafinaz

I Bought A Toy Like Vape Online Mp Tells Dewan Chamber

Tip Mudah Santan Berketul Jadi Elok Semula Letak Sehelai Daun Ini Saja

Unlocking Savings Expert Tips For Finding Discounted Cigarettes In Australia

Kips Bay Decorator Show House Dallas 2024 An Exquisite Showcase Of Design

Empat Filem Rompakan Moden Paling Power Sepanjang Zaman


echo '';
Biodata Terkini Penyanyi Illa Sabry Peserta Gegar Vaganza 2024 Musim 11 GV11

Info Dan Sinopsis Drama Berepisod Senyawa iQIYI Malaysia

Info Dan Sinopsis Drama Berepisod Project Projek Exit Astro Originals

Pelajari Asal Usul Nama Makhluk Seram Kisah Fiksyen Barat Yang Terkenal

10 Fakta Filem Kahar Kapla High Council Yang Ramai Tak Tahu Prekuel Drama Project Projek High Council


Tanam Cili Dari Biji Benih Cili Di Dapur

Empat Sekawan Dipenjara Tiga Tahun Culik Budak Perempuan

Sub Varian Sars Cov 2 Xec 4 November 2024 Africa Cdc

Kkr Aktifkan Bilik Gerakan Bencana Hadapi Monsun Timur Laut

Tan Sri Vincent Tan Dan U Mobile

Cinta Itu Indah Tetapi Duit Lebih Berkuasa